Gas Standards Generator

Gas Standards Generator

MAKER Kin-Tek Laboratories, Inc.
MODEL FlexStream
Application Environmental monitoring meter calibration, Gas analyzer
Product Detail

▷Specification

 

◼ Temperature Control Range: 5 ºC above ambient 20~150ºC (heat only)

◼ Temperature Display Resolution: 0.01 ºC

◼ Standard Flow Range: 0.25~5.0 liter per minute

◼ Optional Flow Ranges: 0.1~0.5, 0.1~1.0, 0.5~10.0 liter per minute

◼ Flow Control over Calibrated Range: ≤ ±1.5 % of reading

◼ Flow Change - 0~Full Scale: <10 sec (2 time constants)

◼ Mode Change - Zero(1 lpm) to Span(1 lpm): < 5 sec (2 time constants)

◼ Output Concentration Range: below 1 ppb to over 1000 ppm

◼ Interfaces: color touch screen display / RS-232, Ethernet

◼ Communication Protocol: Modbus RTU

◼ Power requirement: 115 VAC, 2 A(standard), 230 VAC, 1 A(optional)

◼ Dimensions and weight: 7.5˝ x 13.5˝ x 20˝ (W x H x D). Approximately 30 lbs

◼ [Disposable Permeation tube]

- SO2, H2S, NH3, NO2, CL2, CH3SH, C2H5SH, CCL4, Acrylonitrile, Benzene, Styrene, etc.

    [Refillable Permeation Tubes]

- CO, CO2, NO, N2O, Methane, Ethane, Arsine & Phosphine, Acrylonitrile, benzene, etc.

 

▷Features

 

◼ Complete, integrated, ready-to-use turnkey system

◼ Automated operation with direct readout of concentration TECHNICAL Benefits in units of ppm, ppb or ppt

◼ Modes of Operation: standby, zero, span and purge

◼ Modbus connectivity for integration with existing systems

◼ Flow path suitable for reactive gases - mixture contacts only glass, Teflon® and stainless steel (other materials available)

◼ Accepts disposable permeation tubes, diffusion tubes, ultra-high rate liquid filled tubes, wafer tubes, and prefilled gas fed permeation tubes

◼ Accepts up to 8 disposable tubes with maximum 6 inch length x 1/4 inch diameter (KIN-TEK HRT, SRT and EL tubes)

 

▷Applications

 

Environmental monitoring meter calibration, Gas analyzer