▷Specification |
◼ HCHO
detection technique: Fluorimetric, using Hantzsch (acetyl-acetone) reaction
◼ Linear
detection range: 0.1ppb to 3000ppb (gaseous), 150ng/liter - 5mg/liter
(liquid)
◼ Detection
limit: <100ppt (gaseous), <150ng/liter eq. <2x10^
(-9) molar (liquid)
◼ Sample gas
temperature: >0°C to +120°C
◼ Calibration
and zeroing: Automatic zeroing and semi-automatic calibration using liquid
standards or automatic using internal gas generator (optional)
◼ Weight and
dimension: 20kg, fit for 19" rack (whd: 45cm×15cm×56cm)
◼ Power
requirements: 110VAC / 220VAC, 110W, 24VDC on request
|
▷Features |
◼ Continuous
online monitoring of formaldehyde emissions with unique sensitivity of 100ppt
◼ Analysis of
gaseous and liquid samples with only one instrument
◼ Designed for
climate research, environmental air monitoring and indoor air quality control
◼ Ideal for
emission control of wood-, plastic-, and fabric based products
◼ Can be
employed together with up to 16 emission test chambers
|
▷Applications |
◼ Climate and environmental research / Trace
gas analysis
◼ Online monitoring of HCHO emission /
Emission limit control
◼ Process control / Quality control for the
chemical industry
◼ Chamber testing of wood-, plastic-,
fabric-based products using up to 12 chambers simultaneously
◼ Measurement of indoor air quality
◼ Replacement for the wet chemistry in the
acetyl-acetone method
|