Gas Standards Generator

Gas Standards Generator

MAKER Kin-Tek Laboratories, Inc.
MODEL FlexStream
Application Environmental monitoring meter calibration, Gas analyzer
Product Detail

▷Specification

 

◼ Temperature Control Range: 5 ºC above ambient 20~150ºC (heat only)

 Temperature Display Resolution: 0.01 ºC

 Standard Flow Range: 0.25~5.0 liter per minute

 Optional Flow Ranges: 0.1~0.5, 0.1~1.0, 0.5~10.0 liter per minute

 Flow Control over Calibrated Range: ≤ ±1.5 % of reading

 Flow Change - 0~Full Scale: <10 sec (2 time constants)

 Mode Change - Zero(1 lpm) to Span(1 lpm): < 5 sec (2 time constants)

 Output Concentration Range: below 1 ppb to over 1000 ppm

 Interfaces: color touch screen display / RS-232, Ethernet

 Communication Protocol: Modbus RTU

 Power requirement: 115 VAC, 2 A(standard), 230 VAC, 1 A(optional)

 Dimensions and weight: 7.5˝ x 13.5˝ x 20˝ (W x H x D). Approximately 30 lbs

 [Disposable Permeation tube]

- SO2, H2S, NH3, NO2, CL2, CH3SH, C2H5SH, CCL4, Acrylonitrile, Benzene, Styrene, etc.

    [Refillable Permeation Tubes]

- CO, CO2, NO, N2O, Methane, Ethane, Arsine & Phosphine, Acrylonitrile, benzene, etc.

 

▷Features

 

 Complete, integrated, ready-to-use turnkey system

 Automated operation with direct readout of concentration TECHNICAL Benefits in units of ppm, ppb or ppt

 Modes of Operation: standby, zero, span and purge

 Modbus connectivity for integration with existing systems

 Flow path suitable for reactive gases - mixture contacts only glass, Teflon® and stainless steel (other materials available)

 Accepts disposable permeation tubes, diffusion tubes, ultra-high rate liquid filled tubes, wafer tubes, and prefilled gas fed permeation tubes

 Accepts up to 8 disposable tubes with maximum 6 inch length x 1/4 inch diameter (KIN-TEK HRT, SRT and EL tubes)

 

▷Applications

 

Environmental monitoring meter calibration, Gas analyzer